T1 - Elecsys® Total-Tau and Phospho-Tau (181P) CSF assays: Analytical performance of the novel, fully automated immunoassays for quantification of tau proteins in human cerebrospinal fluid
AU - Lifke, Valeria
AU - Kollmorgen, Gwendlyn
AU - Manuilova, Ekaterina
AU - Oelschlaegel, Tobias
AU - Hillringhaus, Lars
AU - Widmann, Monika
AU - von Arnim, Christine A. F.
AU - Otto, Markus
AU - Christenson, Robert H.
AU - Powers, Jennifer L.
AU - Shaw, Leslie M.
AU - Hansson, Oskar
AU - Doecke, James D.
AU - Li, Qiao-Xin
AU - Teunissen, Charlotte
AU - Tumani, Hayrettin
AU - Blennow, Kaj
PY - 2019/10/1
Y1 - 2019/10/1
N2 - Background: Total tau (tTau) and phosphorylated 181P tau (pTau) are supportive diagnostic cerebrospinal fluid (CSF) biomarkers for Alzheimer's disease. Manual CSF tau assays are limited by lot-to-lot and between-laboratory variability and long incubation/turnaround times. Elecsys® Total-Tau CSF and Phospho-Tau (181P) CSF immunoassays were developed for fully automated cobas e analyzers, allowing broader access in clinical practice and trials. Methods: Analytical performance, reproducibility, method comparisons with commercially available assays, and lot-to-lot and platform comparability (cobas e 601/411) of the Elecsys® CSF assays were assessed. Tau distributions and concentration ranges were evaluated in CSF samples from two clinical cohorts. Results: Both assays showed high sensitivity (limit of quantitation [LoQ]: 63 pg/mL [tTau]; 4 pg/mL [pTau]) and linearity over the measuring range (80–1300 pg/mL; 8–120 pg/mL), which covered the entire concentration range measured in clinical samples. Lot-to-lot and platform comparability demonstrated good consistency (Pearson's r: 0.998; 1.000). Multicenter evaluation coefficients of variation (CVs): repeatability, < 1.8%; intermediate precision, < 2.8%; between-laboratory variability, < 2.7% (both assays); and total reproducibility, < 6.7% (tTau) and < 4.7% (pTau). Elecsys® CSF assays demonstrated good correlation with commercially available tau assays. Conclusions: Elecsys® Total-Tau CSF and Phospho-Tau (181P) CSF assays demonstrate good analytical performance with clinically relevant measuring ranges; data support their use in clinical trials and practice.
AB - Background: Total tau (tTau) and phosphorylated 181P tau (pTau) are supportive diagnostic cerebrospinal fluid (CSF) biomarkers for Alzheimer's disease. Manual CSF tau assays are limited by lot-to-lot and between-laboratory variability and long incubation/turnaround times. Elecsys® Total-Tau CSF and Phospho-Tau (181P) CSF immunoassays were developed for fully automated cobas e analyzers, allowing broader access in clinical practice and trials. Methods: Analytical performance, reproducibility, method comparisons with commercially available assays, and lot-to-lot and platform comparability (cobas e 601/411) of the Elecsys® CSF assays were assessed. Tau distributions and concentration ranges were evaluated in CSF samples from two clinical cohorts. Results: Both assays showed high sensitivity (limit of quantitation [LoQ]: 63 pg/mL [tTau]; 4 pg/mL [pTau]) and linearity over the measuring range (80–1300 pg/mL; 8–120 pg/mL), which covered the entire concentration range measured in clinical samples. Lot-to-lot and platform comparability demonstrated good consistency (Pearson's r: 0.998; 1.000). Multicenter evaluation coefficients of variation (CVs): repeatability, < 1.8%; intermediate precision, < 2.8%; between-laboratory variability, < 2.7% (both assays); and total reproducibility, < 6.7% (tTau) and < 4.7% (pTau). Elecsys® CSF assays demonstrated good correlation with commercially available tau assays. Conclusions: Elecsys® Total-Tau CSF and Phospho-Tau (181P) CSF assays demonstrate good analytical performance with clinically relevant measuring ranges; data support their use in clinical trials and practice.
UR - https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85066258511&origin=inward
UR - https://www.ncbi.nlm.nih.gov/pubmed/31129184
U2 - 10.1016/j.clinbiochem.2019.05.005
DO - 10.1016/j.clinbiochem.2019.05.005
M3 - Article
C2 - 31129184
VL - 72
SP - 30
EP - 38
JO - Clinical Biochemistry
JF - Clinical Biochemistry
SN - 0009-9120
ER -