Guidelines for quality control of PET/CT scans in a multicenter clinical study

Ivalina Hristova, Ronald Boellaard, Paul Galette, Lalitha K. Shankar, Yan Liu, Sigrid Stroobants, Otto S. Hoekstra, Wim J. G. Oyen

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

To date, there is no published detailed checklist with parameters referencing the DICOM tag information with respect to the quality control (QC) of PET/CT scans. The aims of these guidelines are to provide the know-how for effectively controlling the quality of PET/CT scans in multicenter studies, to standardize the QC, to give sponsors and regulatory agencies a basis for justification of the data quality when using standardized uptake values as an imaging biomarker, to document the compliance with the imaging guidelines, to verify the per protocol population versus intent to treat population, and to safeguard the validity of multicenter study conclusions employing standardized uptake value (SUV) as an imaging biomarker which is paramount to the scientific community. Following the proposed guidelines will ensure standardized prospective imaging QC of scans applicable to most studies where SUVs are used as an imaging biomarker. The multitude of factors affecting SUV measurements when not controlled inflicts noise on the data. Decisions on patient management with substantial noise would be devastating to patients, ultimately undermine treatment outcome, and invalidate the utility of SUV as an imaging biomarker usefulness. Strict control of the data quality used for the validation of SUV as an imaging biomarker would ensure trust and reliability of the data.

Original languageEnglish
Article number23
JournalEJNMMI PHYSICS
Volume4
Issue number1
DOIs
Publication statusPublished - 1 Dec 2017

Cite this

Hristova, I., Boellaard, R., Galette, P., Shankar, L. K., Liu, Y., Stroobants, S., ... Oyen, W. J. G. (2017). Guidelines for quality control of PET/CT scans in a multicenter clinical study. EJNMMI PHYSICS, 4(1), [23]. https://doi.org/10.1186/s40658-017-0190-7
Hristova, Ivalina ; Boellaard, Ronald ; Galette, Paul ; Shankar, Lalitha K. ; Liu, Yan ; Stroobants, Sigrid ; Hoekstra, Otto S. ; Oyen, Wim J. G. / Guidelines for quality control of PET/CT scans in a multicenter clinical study. In: EJNMMI PHYSICS. 2017 ; Vol. 4, No. 1.
@article{28d92c2dea744c4d8dff143cd6a03995,
title = "Guidelines for quality control of PET/CT scans in a multicenter clinical study",
abstract = "To date, there is no published detailed checklist with parameters referencing the DICOM tag information with respect to the quality control (QC) of PET/CT scans. The aims of these guidelines are to provide the know-how for effectively controlling the quality of PET/CT scans in multicenter studies, to standardize the QC, to give sponsors and regulatory agencies a basis for justification of the data quality when using standardized uptake values as an imaging biomarker, to document the compliance with the imaging guidelines, to verify the per protocol population versus intent to treat population, and to safeguard the validity of multicenter study conclusions employing standardized uptake value (SUV) as an imaging biomarker which is paramount to the scientific community. Following the proposed guidelines will ensure standardized prospective imaging QC of scans applicable to most studies where SUVs are used as an imaging biomarker. The multitude of factors affecting SUV measurements when not controlled inflicts noise on the data. Decisions on patient management with substantial noise would be devastating to patients, ultimately undermine treatment outcome, and invalidate the utility of SUV as an imaging biomarker usefulness. Strict control of the data quality used for the validation of SUV as an imaging biomarker would ensure trust and reliability of the data.",
author = "Ivalina Hristova and Ronald Boellaard and Paul Galette and Shankar, {Lalitha K.} and Yan Liu and Sigrid Stroobants and Hoekstra, {Otto S.} and Oyen, {Wim J. G.}",
year = "2017",
month = "12",
day = "1",
doi = "10.1186/s40658-017-0190-7",
language = "English",
volume = "4",
journal = "EJNMMI PHYSICS",
issn = "2197-7364",
publisher = "Springer International Publishing AG",
number = "1",

}

Guidelines for quality control of PET/CT scans in a multicenter clinical study. / Hristova, Ivalina; Boellaard, Ronald; Galette, Paul; Shankar, Lalitha K.; Liu, Yan; Stroobants, Sigrid; Hoekstra, Otto S.; Oyen, Wim J. G.

In: EJNMMI PHYSICS, Vol. 4, No. 1, 23, 01.12.2017.

Research output: Contribution to journalArticleAcademicpeer-review

TY - JOUR

T1 - Guidelines for quality control of PET/CT scans in a multicenter clinical study

AU - Hristova, Ivalina

AU - Boellaard, Ronald

AU - Galette, Paul

AU - Shankar, Lalitha K.

AU - Liu, Yan

AU - Stroobants, Sigrid

AU - Hoekstra, Otto S.

AU - Oyen, Wim J. G.

PY - 2017/12/1

Y1 - 2017/12/1

N2 - To date, there is no published detailed checklist with parameters referencing the DICOM tag information with respect to the quality control (QC) of PET/CT scans. The aims of these guidelines are to provide the know-how for effectively controlling the quality of PET/CT scans in multicenter studies, to standardize the QC, to give sponsors and regulatory agencies a basis for justification of the data quality when using standardized uptake values as an imaging biomarker, to document the compliance with the imaging guidelines, to verify the per protocol population versus intent to treat population, and to safeguard the validity of multicenter study conclusions employing standardized uptake value (SUV) as an imaging biomarker which is paramount to the scientific community. Following the proposed guidelines will ensure standardized prospective imaging QC of scans applicable to most studies where SUVs are used as an imaging biomarker. The multitude of factors affecting SUV measurements when not controlled inflicts noise on the data. Decisions on patient management with substantial noise would be devastating to patients, ultimately undermine treatment outcome, and invalidate the utility of SUV as an imaging biomarker usefulness. Strict control of the data quality used for the validation of SUV as an imaging biomarker would ensure trust and reliability of the data.

AB - To date, there is no published detailed checklist with parameters referencing the DICOM tag information with respect to the quality control (QC) of PET/CT scans. The aims of these guidelines are to provide the know-how for effectively controlling the quality of PET/CT scans in multicenter studies, to standardize the QC, to give sponsors and regulatory agencies a basis for justification of the data quality when using standardized uptake values as an imaging biomarker, to document the compliance with the imaging guidelines, to verify the per protocol population versus intent to treat population, and to safeguard the validity of multicenter study conclusions employing standardized uptake value (SUV) as an imaging biomarker which is paramount to the scientific community. Following the proposed guidelines will ensure standardized prospective imaging QC of scans applicable to most studies where SUVs are used as an imaging biomarker. The multitude of factors affecting SUV measurements when not controlled inflicts noise on the data. Decisions on patient management with substantial noise would be devastating to patients, ultimately undermine treatment outcome, and invalidate the utility of SUV as an imaging biomarker usefulness. Strict control of the data quality used for the validation of SUV as an imaging biomarker would ensure trust and reliability of the data.

UR - http://www.scopus.com/inward/record.url?scp=85029691287&partnerID=8YFLogxK

U2 - 10.1186/s40658-017-0190-7

DO - 10.1186/s40658-017-0190-7

M3 - Article

VL - 4

JO - EJNMMI PHYSICS

JF - EJNMMI PHYSICS

SN - 2197-7364

IS - 1

M1 - 23

ER -